Airflow
- Measuring Range: From 40 to 3500 m³/h
- Accuracy: 3 % of the measurement ± 10 m³/h
- Resolution: 1 m³/h
Air Velocity
- Measuring Range: From 0.2 to 10 m/s
- Accuracy: ± 3 % of the measured value ± 0.04 m/s
- Resolution: 0.01 m/s up to 3 m/s and 0.1 m/s above
Temperature (On the Base)
- Measuring Range: From 0° to 50° C
- Accuracy: 2 % of the measured value ± 0.1° C
- Resolution: 0.1° C
Temperature Thermocouple (On the Micromanometer)
- Thermocouple Type: K, J, T, S
- Measuring Range: K : from -200° to 1300° C
- Measuring Range: J : from -100° to 750° C
- Measuring Range: T : from -200° to 400° C
- Measuring Range: S : from 0° to 1760° C
- Accuracy: K, J, T : from -200° to 0° C : ± 0.4° C ± 0.3 % of Reading, from 0° to 1300° C : ± 0.4° C
- Accuracy: S : ± 0.6° C
- Resolution: 0.1° C
Pressure
- Measuring Range: From -2500 to +2500 Pa
- Accuracy: ± 0.2 % of Reading ± 2Pa
- Resolution: From 0.001 to 0.1 Pa
Display
- Large graphic back-lighted LCD / Sizes : 86 x 51 mm
Connectors
- ABS connectors, Ø 7 x 4 mm
Storage Capacity
- 200 datasets of 100 points
Housing
- Shock-proof made of ABS, IP54 protection
Conformity
- Electromagnetical compatibility (NF EN 61326-1 guideline)
Power Supply
- 4 alkaline batteries LR6
- Rechargeable battery Li-ion 3.7 V 4400 mhA (optional)
Auto-Shut-Off
- Adjustable from 0 to 120 min
Languages
- French, English, German, Finnish, Dutch, Italian, Norwegian, Portuguese, Slovak, Swedish, Chinese