8 Inputs Temperature Data Logger – Thermocouples

Price
Request a Quote
Stock
Out of Stock

Specifications

Number of Inputs

  • 8 for HD32.8.8
  • 16 for HD32.8.16

Measuring Range and Accuracy of the Instrument

  • Thermocouple K: -200 to +1370° C, ± 0.1° C up to 600° C / ± 0.2° C over 600° C
  • Thermocouple J: -100 to +750° C, ± 0.1° C up to 400° C / ± 0.2° C over 400° C
  • Thermocouple T: -200 to +400° C, ± 0.1° C
  • Thermocouple N: -200 to +1300° C, ± 0.1° C up to 600° C / ± 0.2° C over 600° C
  • Thermocouple R: +200 to +1480° C, ± 0.3° C
  • Thermocouple S: +200 to +1480° C, ± 0.3° C
  • Thermocouple B: +200 to +1800° C, ± 0.4° C
  • Thermocouple E: 200 to +750° C, ± 0.1° C up to 300° C / ± 0.2° C over 300° C

Resolution

  • 0.05° C (in the range ± 199.99° C)
  • 0.1° C in the remaing range

Drift in Temperature at 20° C

  • 0.02 % / °C

Drift After 1 Year

  • 0.1° C / year

Internal Watch Accuracy

  • 1 min/month max drift

Unit of Measurement

  • °C
  • °F
  • K - Configurable

Memory Capacity

  • (e.g. 1 probe connected = 800,000 acquistions, 8 probes connected = 96,000 acquisitions each probe)
  • Up to 800,000 acqusitions to be divided among all the present inputs max. 64 logging session

Data Logging

  • Instantaneous or delayed, with the possibility of selecting the storage start and end time.

Storage Interval Selection

  • 2, 5, 10, 15, 30 seconds
  • 1, 2, 5, 10, 15, 20, 30 minutes.
  • 1 hour

Data Download

  • RS-232C from 1200 to 38,400 baud, galvanically isolated.
  • USB 1.1 - 2.0 galvanically isolated.
  • Sub D 9-Pole male connector.

Security of Stored Data

  • Unlimited

Power Supply

  • 4 per 1.5 V alkaline batteries type C-BABY External 12 Vdc-1A power supply
  • Connector, external Ø 5.5 mm, internal Ø 2.1 mm
  • Power supply via the PC USB port.

Current Consumption at 6 Vdc

  • < 60 μA when the instrumen is of
  • < 60 μA in sleep mode with all probes connected
  • < 40 mA during data logging with all probes connected

Voltage Insulation

  • 60 V between inputs
  • 500 V between inputs and power supply

Autonomy

  • 200 hours with 7800 mAh alkaline batteries and all probes connected

Connection

  • Miniature female socket for thermocouple

Operating Temperature

  • -5° to 50° C

Storage Temperature

  • -25° to 65° C

Working Relative Humidity

  • 0 to 90 %RH, no condensation

Protection Degree

  • IP64

Dimensions (Length x Width x Height)

  • 220 x 180 x 50 mm

Weight

  • 1100 g (complete with batteries)

Materials

  • ABS
  • Polycarbonate
  • Aluminum

Display

  • Backlit graphic LCD 128x64 pixel

Keyboard

  • 15 keys, configurable also without PC.
  • Security password for keyboard locking

show more