Price
Request a Quote
Stock
Out of Stock

  • Transmitter - Standard

Specifications

Accuracy

  • ± 5 µA
  • Basic Measurement Accuracy: ± 0.01 % of Full Range Input (FRI) ± 0.05 % of Reading
  • Basic Measurement Accuracy: ± 0.04 % of FRI ± 0.04 % of Reading or 0.5° C (0.9° F), whichever is greater
  • Basic Measurement Accuracy: ± 10 µV ± 0.07 % of Reading
  • Basic Measurement Accuracy: 0.1 %
  • Cold Junction Error: ± 0.5° C (± 0.9° F)

Approvals & Certifications

  • ATEX: II 1 G EEx ia IIC T4-T6
  • EMC: BS EN 61326
  • FM: IS/I/1/ABCD/T4

Cable Length

  • 3,281 feet (1,000 metres) - Max.

Communication

  • PC Interface: USB Connector

Control Functions Characteristics

  • Linear

Current

  • Excitation Current: 300-500 µA

Enclosure, Body Material

  • Noryl

Humidity

  • Ambient Humidity: 10 to 90 % Relative Humidity, Non-Condensing
  • I.S Version: 0 to 100 % Relative Humidity

Input

  • 3-Wire Potentiometer
  • Voltage Source

Input Impedance

  • 10 megaohms

Isolation

  • Input/Output Breakdown Isolation: 500 Vac RMS

Linearization

  • RTD: BS 1904
  • RTD: BS EN 60751
  • RTD: DIN 43760
  • RTD: JISC 1604
  • Thermocouple: BS 4937
  • Thermocouple: IEC 584-3

Operating Temperature

  • Ambient Operating Range & Cold Junction Range: -40° to 85° C (-40° to 185° F)

Output

  • 20.2 mA
  • 23 mA - Max.

Output Load

  • 700 ohms at 24 V, Max.

Power Supply Effect

  • Voltage Effect: 0.2 μA/V

Range

  • -10 to 75 mV

Resistance

  • 10 to 390 ohms
  • Maximum

Response Time

  • Filter OFF: < 1 second

RTD Type

  • Pt100

Sensor Protection

  • Burn-Out

Sensor, Probe Type

  • E Thermocouple
  • J Thermocouple
  • K Thermocouple
  • L Thermocouple
  • N Thermocouple
  • RTD
  • R Thermocouple
  • S Thermocouple
  • T Thermocouple

Span - Minimum

  • 25° to 100° C (77° to 212° F)
  • 5 mV
  • 5%

Stability

  • 0.1 % of Full Range Input or 0.1° C (0.18° F)/Year

Storage Temperature

  • Ambient Storage Temperature: -50° to 100° C (-58° to 212° F)

Supply Voltage

  • 10-35 V

Temperature Effect

  • Cold Junction Tracking: 0.05° C/°C
  • Lead Resistance Effect: 0.002° C/ohm

Temperature Range

  • Sensor: -200° to 1,760° C (-328° to 3,200° F)
  • Sensor: -200° to 850° C (-328° to 1,562° F)

Thermal Drift

  • Span: 0.01 %/°C
  • Zero: 0.008° C/°C
  • Zero: 0.1 µV/°C

Update Time

  • 250 milliseconds, Max.
  • Warm-Up: 120 seconds to Full Accuracy
show more