LP471UVBC – UV-BC Irradiance Probe

Brand: Delta OHM S.R.L.
Does not ship to Ohio, United States

LP471UVBC is a radiometric probe for measuring irradiance (W/m²) in the 210 to 355 nm spectral range with peak at 265 nm. It is equipped with a connector with SICRAM module and a 2 m cable.

The probe consists of a photodiode with an appropriate filter and is provided with a diffuser for correct measurement according to the cosine law.

It is typically used in the verification of germicidal lamps used in sterilization systems working with LEDs as UV light sources.

The probe is available as well already combined in ready to use kits. Depending on your needs, you can choose a basic one composed by the probe and our HD2302.0 luxmeter to simply measure and read the values (UVBC-CK02-ACC).

For a more advanced version with data logging feature, the probe is combined with our HD2102.2 portable luxmeter datalogger (UVBC-CK01-ACC).

For both solutions, the correctness of your measurements are also assured by the ISO 17025 Calibration Certificate issued in our own accredited Photo-radiometry laboratory.

Specifications

Measuring Range

  • 1 ∙ 10-³ to 2000 W/m²

Spectral Range

  • 210 to 355 nm (Peak 265 nm)

Calibration Uncertainty

  • < 7 % (calibration @ 254 nm)

f3 (Linearity)

  • < 2 %

f5 (Fatigue)

  • < 0.5 %

Drift After 1 Year

  • < 2 %

Working Temperature

  • 0° to 50° C

Connection to the Instrument

  • Connector with SICRAM module and 2 m cable

  • Verification of germicidal lamps used in sterilization systems that use LEDs as UV light sources.

Documentation

show more