SEM203TC Temperature Transmitter
Brand: Status InstrumentsSEM203TC Temperature Transmitter
Brand: Status InstrumentsThe SEM203TC is a cost effective “smart” in head transmitter that accepts thermocouple temperature sensors and converts sensor output over a configured range to a standard industrial (4-20 mA) transmission signal.
A simple push button operation allows the user to select TC type, Burnout direction, Select fixed ranges and trim (4 and 20) mA points.
The SEM203TC in head transmitter incorporates the latest digital technology to ensure accurate drift free performance.
If required the desired range can be specified at the time of order, removing the need for user configuration.
If the range is not specified then the transmitter will be shipped with the default range of (0 to 1,000)° C type K.
Specifications
Brand
Accuracy
- ± 0.02 % of Full Scale - for Range mV
- ± 0.1 % of Full Scale ± 0.5° C (± 0.9° F) - for Thermocouple Typical K, J, E, N, R & S
- ± 0.2 % of Full Scale ± 0.5° C (± 0.9° F) - for Thermocouple Typical T
- 5 µA or (mA Output/2,000), whichever is greater
- Cold Junction: ± 0.5° C (± 0.9° F) Tracking ± 0.05° C/°C at -40° to 85° C (-40° to 185° F)
Approvals & Certifications
- CE
- EMC: BS EN 61326
- IEC 61000-4-2, 61000-4-3, 61000-4-4, 61000-4-5
Configuration
- Push Button
Electrical Connectors
- Screw Terminal
Humidity
- 10 to 90 % Relative Humidity, Non-Condensing
Indication
- LED
Isolation
- 250 Vdc
Operating Temperature
- -40° to 85° C (-40° to 185° F)
Output
- 4-20 mA Current Loop, 2-Wire
- Maximum
- Minimum: 3.8 mA (In Low Burn-Out Condition)
Output Load
- Maximum
Power Supply
- 12-30 Vdc
Power Supply Effect
- ± 0.2 μA/V - Loop Voltage Effect
Response Time
- 1 second
Sensor Protection
- Sensor Burn-Out (Either Up or Down Scale Output)
Sensor, Probe Connection
- 2-Wire
Sensor, Probe Type
- E Thermocouple
- J Thermocouple
- K Thermocouple
- N Thermocouple
- R Thermocouple
- S Thermocouple
- T Thermocouple
Stability
- ± 0.1 % of Reading/°C
- ± 0.1° C/°C at Span
- ± 0.15° C/°C at 0° C
Storage Temperature
- -50° to 90° C (-58° to 194° F)
Temperature Range
- -200° to 1,760° C (-328° to 3,200° F)
Thermal Drift
- ± 2 μA/°C
Turn-On Time
- Start Up Time: within 8 seconds (Output < 4 mA During Start Up)
Update Time
- 1 Minute to Full Accuracy
- 500 milliseconds