Radiometric probe for measuring the IRRADIANCE in the UVC spectral range 220 nm to 280 nm, peak at 260 nm, quartz diffuser for cosine correction.

Measuring range
  • 1 mW/m² to 2000 W/m².
Cable length 2 m.

Specifications

Measuring Range: Resolution (W/m²)

  • 1.0·10-3 to 999.9·10-3 : 0.1·10-3
  • 1.000 to 19.999 : 0.001
  • 20.00 to 199.99 : 0.01
  • 200.0 to 1999.9 : 0.1

Spectral Range

  • 220 nm to 280 nm (Peak 260 nm)

Calibration Uncertainty

  • < 5 %

f3 (Linearity)

  • < 1 %

f4 (Instrument Reading Error)

  • ± 1digit

f5 (Fatigue)

  • < 0.5 %

Drift After 1 Year

  • < 2 %

Working Temperature

  • 0° to 50° C

Documentation

show more