HD2178 – Thermocouple – Pt100 Thermometer Data Logger

Brand: Delta OHM S.R.L.
Does not ship to United States

HD2178.1 and HD2178.2 are portable instruments with a large LCD display.

These instruments measure temperature by means of immersion, penetration, contact or air probes with Pt100, Pt1000 or thermocouple probes.

You can connect a 3 or 4 wires Pt100 sensor or a 2 wires Pt1000 sensor to B input, a K, J, T, N, E type thermocouple to input A.

Probes to B input, a 8-poles DIN45326 connector, are equipped with an automatic detection module, with the factory calibration settings already being memorized inside.

A input is equipped with a miniature female polarized connector for thermocouple probes.

The instrument HD2178.2 is a data logger; it stores up to 80.000 samples that can be transferred to a PC when connected to the instrument through a RS232C serial port or a USB 2.0 port.

It is possible to configure the storage interval, the printing and the baud rate or to a portable printer by the menu.

Functions Max, Min and Avg calculate maximum, minimum and average values.

Further functions are: REL relative measure, HOLD and automatic switching-off system (excludable).

Instruments have IP66 protection degree.

Specifications

Pt100 Measuring Range

  • -200…+650° C

Pt1000 Measuring Range

  • -200…+650° C

Accuracy

  • ± 0.05° C

Measuring Range

  • K: -200 to +1370° C
  • J: -100 to +750° C
  • T: -200 to +400° C
  • N: -200 to +1300° C
  • E: -200 to +750° C

Instrument Accuracry

  • Thermocouple K: ± 0.1° C up to 600° C, ± 0.2° C over 600° C
  • Thermocouple J: ± 0.1° C up to 400° C, ± 0.2° C over 400° C
  • Thermocouple T: ± 0.1° C
  • Thermocouple N: ± 0.1° C up to 600° C ± 0.2° C over 600° C
  • Thermocouple E: ± 0.1° C up to 300° C ± , 0.2° C over 300° C

Resolution

  • 0.1° C

Temperature Drift at 20° C

  • 0.02 %/°C

Drift After 1 Year

  • 0.1° C/year

Unit of Measurement

  • °C - °F

Measured Values Storage - Model HD2178.2 Type

  • 2000 pages each one containing 40 samples

Quantity

  • 80000 samples in total

Storage Interval Can be Selected Among

  • 1,5,10,15,30 s,
  • 1,2,5,10,15,20,30 min
  • 1 hour

Security of Stored Data

  • Unlimited, independent of battery charge conditions

Batteries

  • Four 1.5 V type AA batteries

Autonomy

  • 200 hours with 1800 mAh alkaline batteries

Current Consumption with Instrument Off

  • 20 μA

Mains

  • 12 Vdc / 1000 mA Output mains adapter

Serial Interface RS-232C Type

  • RS-232C galvanically isolated

Baud Rate

  • can be set up from 1200 to 38400 baud

Data Bit

  • 8

Parity

  • None

Stop Bit

  • 1

Flow Control

  • Xon/Xoff

Serial Cable Length

  • Maximum 15 m

Print Interval

  • 1,5,10,15,30 s,
  • 1,2,5,10,15,20,30 min
  • 1 hour

USB Interface - Model HD2178.2 Type

  • 1.1 - 2.0 galvanically isolated

Input for RTD Probes

  • 8 pole Male DIN 45326 connector

Input for TC Probes

  • 2-Pole female polarized standard miniature Connector

RS-232C Serial Interface

  • 8-Pole MiniDin connector

USB Interface

  • Type B MiniUSB connector

Mains Adapter

  • 2-Pole Connector (positive at centre)

Operating Temperature

  • -5...50° C

Storage Temperature

  • -25 … 65° C

Working Relative Humidity

  • 0 … 90 % RH, no condensation

Protection Degree

  • IP66

Dimensions (Length x Width x Height)

  • 185 x 90 x 40 mm

Weight

  • 470 g (complete with batteries)

Materials

  • ABS, rubber

Display

  • 2 rows 4-1/2 digits plus symbols Visible area: 52 x 42 mm

Date and Time

  • In real time

Time Accuracy

  • 1 min/month max drift
show more